A.
Take a back-up driving unit to the bedside
B.
Restart the driving unit
C.
Clamp the return line
D.
Re-insert the pumphead in the primary motor
E.
Restart the pump; as minimum RPMs are reached to prevent backflow (depending from IFU insert), remove the last clamp on the return line
F.
Contact technical assistance to thoroughly assess the failing driving unit
G.
Report the issue to the manufacturer as needed
H.
Strictly monitor the circuit and the patient
I.
Document the event and related procedures in the patient's chart
J.
If applicable, report the incident in the ELSO registry
This concludes No Flow.